5Y.15.PM Precision Balance
Laboratory Precision Under Industrial Conditions
- The ability to weigh loads up to 120 kg on a large weighing pan with high accuracy and good repeatability,
- minimum eccentricity error over the entire range,
- quick and stable measurement (achievable within 1 second),
- parameter stability over time and during transport,
- robust design based on a larger and taller ,
- built-in 4-point overload protection system,
- weighing pan dimensions adapted to the accuracy of the balance,
- internal adjustment,
are the distinguishing features of RADWAG PM balances. All of this is possible thanks to the use of a cutting-edge measuring system based on enhanced MONOBLOCK® technology. The change in the design of the MONOBLOCK® has made it possible to work with a readability of 0.01 g at a maximum capacity of 20 kg and 0.2 g at a maximum capacity of 120 kg, with a repeatability of 0.01 g, and resistance to changes in ambient conditions.
5Y.PM balances also include:
- 10-inch touch screen,
- easy start-up, intuitive operation, and ergonomics,
- compliance with 21 CFR Part 11,
- Digital Weighing Auditor,
- multi-step user verification,
- Ambient Light alerts,
- Hotspot,
- RFID,
- Live Note text notes,
- handy measurement history within the Handy Library,
- widgets.
Application
The 5Y.PM balances are more accurate, faster, and more versatile than their predecessors and therefore meet the requirements for equipment used in companies representing various industries. Dosing, formulation-making, and parts counting are just some of the possible applications of these devices. Wherever their potential for impressive accuracy and high load capacity is to be realised, PM balances will prove themselves.
21 CFR Part 11
Do you work in the pharmaceutical industry? Do you need a digital signature?
We are offering the laboratory balance which as a standalone fully meets the requirements of 21 CFR Part 11/ EU GMP Annex 11.
- Full data security
- Electronic signatures
- Validations
- Electronic records
- Audit trails
Modbus TCP/IP
is one of the most popular communication protocols in industrial automation. Implementing it in the 5Y balances and CY10 scales makes the devices of this series friendly to automation integrators.
A balance or scale in a Modbus system is a slave device and performs tasks specified by the master device which is most often a PLC. Protocol Modbus TCP/IP in the 5Y balances and C10 scales enables, among other things, the realization of mass readout, zeroing and tare functions of the device, as well as tare settings and MIN and MAX thresholds.
Under-pan weighing
Weighing instruments purchased after 15th February 2023 will allow under-pan weighing at any time after the purchase. All you need to do is just to order a dedicated hook.
THBR 2.0 System
Cooperation with THBR 2.0 System for ambient conditions monitoring.
Datasheet
Metrological Parameters | |
---|---|
Maximum capacity [Max] | 15 kg |
Minimum load | - |
Readability [d] | 0.01 g |
Verification unit [e] | - |
Tare range | -15 kg |
Standard repeatability [5% Max] | 0.004 g |
Standard repeatability [Max] | 0.012 g |
Minimum weight (USP) | 8.2 g |
Minimum weight (U=1%,k=2) | 0.82 g |
Linearity | ±0.03 g |
Stabilization time | 1 s |
Adjustment | Internal (automatic) |
OIML Class | - |
Physical Parameters | |
Leveling system | Semi-automatic – LevelSENSING |
Display | 10″ graphic colour touchscreen |
Delivery components | Precision Balance, weighing pan, fabric dust cover, power supply. |
Weighing pan dimensions | 200×185 mm |
Packaging dimensions (W x D x H) | 520×520×280 mm |
Net weight | 12 kg |
Gross weight | 14.2 kg |
Construction | |
Protection class | IP 43 |
Communication Interface | |
Interfaces | 2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot |
Electrical Parameters | |
Power supply | Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A Balance: 12 – 15V DC 1.6A max; 10–19W* |
Environmental Conditions | |
Operating temperature | +10 – +40 °C |
Wi-Fi® is a registered trademark of Wi-Fi Alliance®.
Repeatability is expressed as a standard deviation from 10 cycles of mass standard weighing.
Stabilization time dependson the ambient conditions and the dynamics of weighing pan loading; specified for FAST profile.
*Power consumption depends on the terminal configuration and the number and type of external devices connected.
Quality / Certification Assurance with every Solution
When you work in an industry that involves transporting multiple vehicles with heavy units and materials, weighing becomes a necessity. Driving a vehicle on the road is not road legal due to its weight. Having your vehicles weighed incorrectly could ultimately leave your business vulnerable to prosecution.