Maximum number of verification intervals: nLC = 3000
Combined error 0,03% of Full Scale Output (F.S.)
Non-Repeatability: 0,01% F.S.
Full Scale Output: 3mV/V +/- 0,08%
Temperature effect on zero: 0,017% /°C
Temperature effect on full scale output: 0,014% /°C
Compensated Temperature Range: -10°C/+40°C
Operating Temperature Range: -35°C/+65°C
Creep error after 30 minutes: 0,02% F.S.
Maximum tolerated excitation voltage: 18 VDC
Input Resistance: 430 ± 60 Ohm
Output Resistance: 350 ± 3.5 Ohm
Zero balance: +/-1% F.S.
Insulation Resistance: >5000 MOhm
Safe Overload: 150% F.S.
Ultimate Overload: 300% F.S.
4-wire shielded cable, 6m long, 5 mm diameter
ATEX VERSION
The CCATEX certificate makes the load cell suitable for use in hazardous areas with protection methods according to: ATEX II 1G Ex ia IIC T6 (Ta -20÷+40°C) TX (Ta -20÷+65°C) Ga ATEX II 1D Ex ta IIIC TX°C (Ta -20÷+40°C) TX°C (Ta -20÷+65°C) Da IP65. The certificate includes the manual and the ATEX CE conformity declaration (available in EN, DE, FR, and IT).
IMPORTANT: the certificate refers to the single load cell. For systems with more load cells, it is necessary to order one certificate for each load cell.